GroLine Monitor for Hydroponic Nutrients with Inline Probe
Hanna Instruments’ GroLine Monitor for Hydroponic Nutrients with Inline Probe makes nutrient management easier with 24/7 pH&EC measurements for consistent yields every time, a Good Laboratory Practice (GLP) feature for professional accuracy, and hassle free calibration of pH and EC with a single Quick Cal solution. . This easy to setup, simple to use monitor was designed with hydroponics, aquaponics, and greenhouses in mind.
Readings at a Glance
A large clear LCD displays all of your essential readings so you know what your hydroponic nutrients solution is doing 24/7. High and low alarms flash warning you if something needs adjusting.
Monitor Changes Over Time
The GroLine Monitor automatically logs your results every 15 minutes over the last 30 days. Minimum, maximum, and average values are stored so you can recognize when patterns arise and help prevent future problems.
One Probe - Three Measurements
This unit's 3-in-1 pre-amplified inline probe measures pH, EC/TDS, and temperature in a single, durable, polypropylene body making the GroLine Monitor easy to handle and easy to set up and it has a 3/4” threaded connection for insertion into a pipe for flow through monitoring.
Quick Cal for Less Maintenance
A single calibration solution for pH and conductivity makes calibration fast and ensures all of your measurements are accurate all of the time for better plant health and better yield.
Professional Level Accuracy
Advanced algorithms and greater pH and TDS/ppm accuracy ensure operational consistency and a finer understanding of your plant’s nutritional needs.
Professional Measurements
The GroLine Monitor is made by growers and scientists who are committed to helping you take your growing operation to the next level.
Calibration
pH calibration at one or two-point calibration (using pH 4.01,7.01, 10.01 buffers) using auto buffer recognition; one-point calibration using quick calibration solution; EC calibration at one-point at 1.41 mS/cm or 5.00 mS/cm using auto standard recognition; one-point calibration using quick calibration solution
Product Features
- Three sensors combined in a single rugged probe body made for continuous immersion up to 2 meters: pH electrode with renewable cloth junction, amperometric EC/TDS sensor, and an internal temperature sensor for temperature compensated readings
- IP65 rated enclosure: designed to withstand harsh grow environments
- Wide pH Range: measure pH from 0.00 to 14.00 pH
- Selectable EC to TDS/ppm conversion factor: choice of either a 0.5 or 0.7 conversion factor
- Automatic Temperature Compensation: all readings are compensated for variations in temperature, temperature displayed in oC or oF along with pH, EC or TDS reading
- Large LCD with plant-friendly green backlighting: LCD can display all three essential hydroponic nutrient measurements at one time, a 3-second cycle of single measurements, or a real-time graph screen with options for measurement selection and log recall
- Data logging for 30 days: logs every 15 minutes for last 30 days, stores min, max, and average values
- Data transfer: USB-C port for easy data transfer to memory stick or PC
- High and low alarms: warns when your nutrient solution is out of desired range by flashing display and message
- Ambient light sensor: for automatic LCD dimming
GroLine Monitor Specifications
-
pH Range 0.00 to 14.00 pH
-
pH Resolution 0.01 pH
-
pH Accuracy (@25°C/77°F) ±0.05 pH
-
Temperature Range 0.0 to 60.0°C / 32.0 to 140.0°F
-
Temperature Resolution 0.1°C / 0.1°F
-
Temperature Accuracy (@25°C/77°F) ±0.5°C /±1°F
-
EC Range 0.00 to 10.00 mS/cm
-
EC Resolution 0.01 mS/cm
-
EC Accuracy (@25°C/77°F) ±0.1mS/cm (0.00 to 5.00mS/cm); ±0.2mS/cm (5.00 to 10.00mS/cm)
-
TDS Range 0 to 5000 ppm (TDS conversion factor 0.5 (500 CF)); 0 to 7000 ppm (TDS conversion factor 0.7 (700 CF))
-
TDS Resolution 10 ppm
-
TDS Accuracy (@25°C/77°F) ±2% F.S.
-
Temperature Compensation pH: automatic (0.0 to 60.0ºC/ 32.0 to 140.0°); EC: automatic with β set at 1.9%/°C
-
Log Feature Automatic logging for last 30 days at 15 minute interval
-
Alarms High and Low with enable/disable option for all parameters
-
Connectivity USB Type-C
-
GLP Good Laboratory Practice with last 5 pH and EC calibration history
-
Power Supply 12VDC adapter (included)
-
Environment -5 to 50°C (23 to 122°F); RH max 95% non-condensing
-
Dimensions 125 x 185 x 38 mm (4.92 x 7.28 x 1.49")
-
Weight 333g ( 11.7oz.)